Pipeline gas supplies are connected and the pressures are
correct.
•
Cylinder valves are closed.
•
Models with cylinder supplies have a cylinder wrench attached
to the system.
•
Models with cylinder supplies have a reserve supply of O2
connected to the machine during system checkout.
•
The necessary emergency equipment is available and in good
condition.
•
Equipment for airway maintenance, manual ventilation, tracheal
intubation, and IV administration is available and in good
condition. In the case of system failure, the lack of immediate
access to alternative means of ventilation can result in patient
injury.
•
Applicable anesthetic and emergency drugs are available.
•
Check that the brake is set to prevent movement.
•
The power cord is connected to an electrical outlet. The mains
indicator comes on when AC power is connected. If the indicator
is not on, the system does not have mains (electrical) power.
Use a different outlet, close the circuit breaker, or replace or
connect the power cable.
•
If an optional suction regulator is present, ensure there is
adequate suction.
•
If an optional O2 flowmeter is present, ensure there is adequate
flow.
Electrical safety
Do not connect non-medical electrical equipment directly to the AC
outlet at the wall instead of an AC power source which uses a
separating transformer. Doing so may increase enclosure leakage
current above levels allowed by IEC 60601-1 in normal conditions
and under single-fault conditions. This may cause an unsafe
electrical shock to the patient or operator.
After connecting anything to these outlets, conduct a complete
system leakage current test (according to IEC 60601-1).
The system provides connections for items such as
printers, visual displays and hospital information
networks (only connect items that are intended to be part
of the system). When these items (non-medical
equipment) are combined with the system, these
precautions must be followed:
•
Do not place items not approved to IEC 60601-1
closer than 1.5 m to the patient.
•
All items (medical electrical equipment or non-
medical electrical equipment) connected to the
system by a signal input/signal output cable must be
WARNING
Aisys CS²
1-18
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...