Ventilation modes factory default settings
The ventilation mode factory default settings for the ADULT case
default show in the following table. An * indicates the setting is not
used for the ventilation mode. See the "
" section for
all other case default values.
.
Setting
Range
Mode
VCV
PCV
PCV-VG SIMV VCV SIMV PCV SIMV
PCV-VG
PSVPro CPAP
+PSV
TV
20-1500 ml
500
*
500
500
*
500
*
*
RR
4-100 bpm
12
12
12
*
*
*
*
*
RR
2-60 bpm
*
*
*
12
12
12
12
*
Minimum RR
4-60 /min
*
*
*
*
*
*
*
12
I:E
2:1 - 1:8
1:2
1:2
1:2
*
*
*
*
*
Tpause
Off, 5-60% of
Tinsp
Off
*
*
Off
*
*
*
*
PEEP
Off, 4-30
cmH2O
Off
Off
Off
Off
Off
Off
Off
Off
Pmax
12-100
cmH2O
40
40
40
40
40
40
40
40
Pinsp
5-60 cmH2O
*
5
*
*
5
*
5
5
Psupport
Off, 2-40
cmH2O
*
*
*
2
2
2
2
2
Tinsp
0.2-5 seconds *
*
*
1.7
1.7
1.7
1.7
1.7
Rise Rate
Auto, 1-10
*
Auto
*
Auto
Auto
*
Auto
Auto
Rise Rate
1-10
*
*
5
*
*
5
*
*
PSV Rise Rate Auto, 1-10
*
*
*
*
*
Auto
*
*
Trig Window
Off, 5-80% of
Texp
*
*
*
25
25
25
25
*
Flow Trigger
0.2-10 l/min
*
*
*
2.0
2.0
2.0
2.0
2.0
End of Breath
5-75% of peak
inspiratory
flow
*
*
*
30
30
30
30
30
Backup Time
10-30 seconds *
*
*
*
*
*
30
*
Exit Backup
Off, 1-5
spontaneous
breaths
*
*
*
*
*
*
2
*
11 Specifications and theory of operation
2067226-001
11-35
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...