Super user mode
Do not enter Super User mode when a patient is
connected to the system. Gas flow will cease, and the
system must be powered down in order to restart gas
flow.
Changes made in Super User mode affect the system
configuration. All changes made are permanent and
preserved until changed again.
Several settings can be changed in Super User mode. These
changes should only be made by the person responsible for the
configuration of the system.
To access the Super User mode, select
System Setup
and then
Super User
. The Super User mode is password protected. Contact
the sales representative to obtain the password.
The system must be turned off to exit Super User mode.
Menu items are dependant on system configuration. Inactive menu
items are gray and are not selectable.
The menu selections shown in the examples are the factory default
values. The available settings are shown to the right of the menu.
WARNING
•
Note
Note
Aisys CS²
12-2
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...