Alarms On Off
Use the
Alarms On/Off
menu to turn On or turn Off the
CO2 Alarms
Limits
and
MV/TV Alarm Limits
during manual ventilation. Alarm
limits are enabled at the start of mechanical ventilation.
The CO2 ‘Apnea’ alarm cannot be turned off through the
Alarms
On/Off
menu. Only the ‘EtCO2’ and ‘FiCO2’ alarms are turned off
using this menu.
Disable alarm limits
1.
Select
Alarms On/Off
.
2.
View the alarm limit status.
3.
Select
Confirm
to change the alarm limit settings. Select
Cancel
to leave the alarm limits unchanged.
•
If the
CO2 Alarms Limits
and
MV/TV Alarm Limits
are
On
, the alarm limits change to
Off
.
•
If the
CO2 Alarms Limits
or
MV/TV Alarm Limits
are
On
,
the alarm limits change to
Off
.
•
If the
CO2 Alarms Limits
and
MV/TV Alarm Limits
are
Off
, the alarm limits change to
On
.
Note
3 Operation
2067226-001
3-21
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...