Typical performance
.
M- and E-series
CARESCAPE*
CO2
Measurement range 0 to 15 vol% (0 to 15 kPa,
0 to 113 mmHg).
Measurement rise time less than 400 ms
typical.
Ac/- (0.2 vol% + 2% of reading).
6 hour drift less than 0.1 vol%
Gas cross effects less than 0.2 vol% (O2,
N2O, anesthetic agents).
Measurement range 0 to 15 vol% (0 to 15 kPa,
0 to 113 mmHg).
Measurement rise time less than 260 ms
typical.
Ac/- (0.2 vol% + 2% of reading).
6 hour drift less than 0.1 vol%
Gas cross effects less than 0.2 vol% (O2,
N2O, anesthetic agents).
O2
Measurement range 0 to 100 vol%.
Measurement rise time less than 400 ms
typical.
Ac/- (1 vol% + 2% of reading).
6 hour drift less than 0.2 vol%
Gas cross effects less than 1 vol% anesthetic
agents, less than 2 vol% N2O.
Measurement range 0 to 100 vol%.
Measurement rise time less than 260 ms
typical.
Ac/- (1 vol% + 2% of reading).
6 hour drift less than 0.3 vol%
Gas cross effects less than 1 vol% anesthetic
agents, less than 2 vol% N2O.
N2O
Measurement range 0 to 100 vol%.
Measurement rise time less than 450 ms
typical.
Ac/- (2 vol% + 2% of reading).
6 hour drift less than 0.1 vol%
Gas cross effects less than 2 vol% anesthetic
agents.
Measurement range 0 to 100 vol%.
Measurement rise time less than 320 ms
typical.
Ac/- (2 vol% + 2% of reading).
6 hour drift less than 0.3 vol%
Gas cross effects less than 2 vol% anesthetic
agents.
Anesthetic agents
Measurement range Hal, Enf, Iso 0 to 6% vol
%.
Measurement range Sev 0 to 8 vol%.
Measurement range Des 0 to 20 vol%.
Measurement rise time less than 400 ms
typical.
Ac/- (0.15 vol% + 5% of reading).
6 hour drift for Hal, Enf, Iso, Sev less than 0.1
vol%
6 hour drift for Des less than 0.2 vol%
Gas cross effects less than 0.15 vol% N2O.
Measurement range Hal, Enf, Iso 0 to 6% vol
%.
Measurement range Sev 0 to 8 vol%.
Measurement range Des 0 to 20 vol%.
Measurement rise time Des, Enf, Iso, Sev less
than 420 ms typical.
Measurement rise time Hal less than 800 ms
typical.
Ac/- (0.15 vol% + 5% of reading).
6 hour drift for Hal, Enf, Iso, Sev less than 0.1
vol%
6 hour drift for Des less than 0.3 vol%
Gas cross effects less than 0.15 vol% N2O.
*The effects caused by N2O to the measurement of CO2, O2, and anesthetic agents are automatically
compensated for. The effects caused by anesthetic agents to the measurement of CO2 and N2O are automatically
compensated for.
11 Specifications and theory of operation
2067226-001
11-21
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...