2.
Select the item to change.
Set
VCV Cardiac Bypass
to
Yes
to enable alveolar support
during cardiac bypass. This menu item is only available on
systems with the VCV Cardiac Bypass option installed.
Set
Circuit Compliance
to
On
to allow tidal volume
compensation for circuit compliance.
Set
Sample Compensation
to
On
to allow tidal volume
compensation for gas module sampling.
Set the
TV for Ideal Body Weight
value to compute RR and TV
on a case.
3.
Select
Back
to return to the
System Config.
menu.
Alarm settings
Use the
Alarm Settings
menu to set
External Gas Monitor
,
Apnea
Alarm Filter
, and
Show Alarm Limits
default settings.
Setting alarm defaults
1.
From Super User mode, select
System Config.
and then
Alarm
Settings
.
2.
Select the item to change.
Set
External Gas Monitor
to
Yes
only if the system uses a
stand-alone monitor for O2, agent, and CO2. Selecting
Yes
disables the O2 limit alarms and the ‘O2 monitoring not
connected’ alarm when there is no O2 cell or airway module
detected in the system. The ‘AA, CO2 monitoring not connected’
alarm is also disabled.
Set
Apnea Alarm Filter
to
Enable
to use CO2 breaths to filter
the volume apnea alarms when patient weight is set to the
minimum setting. ‘Volume Apnea Off’ shows in the message
field when the apnea alarm filter conditions are met. Set
Apnea
Alarm Filter
to
Disable
to shut off the use of CO2 breaths for
filtering volume apnea alarms.
Set
Show Alarm Limits
to
Yes
to show the alarm limits for the
primary parameter in the numeric and digit fields.
3.
Select
Back
to return to the
System Config.
menu.
Parameter settings
Use the
Parameter Settings
menu to set
Principal Volume
,
Principal O2
,
Principal Agent
,
TV Based on
,
CO2 Numbers
, and
Insp Flow
default settings.
Aisys CS²
12-6
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...