Airway modules
The optional compact airway modules measure and monitor gases
delivered to the patient and exhaled through the breathing circuit.
The modules consist of:
•
An infrared sensor for measuring CO2, N2O, and anesthetic
agents.
•
A paramagnetic O2 sensor.
•
A gas sampling system with D-fend water separation system.
Systems with both an airway module and an O2 cell will display the
patient inspired O2 value obtained from the airway module.
Respiratory rate is the frequency of peak (end tidal) CO2
measurements per minute. A breath is defined as a change in the
CO2 signal that exceeds 1% (8 mmHg). All concentrations are
measured and displayed breath by breath.
Remove the airway sampling line for the patient’s airway
and seal the sample port while nebulized medications are
being delivered. Nebulized medications interfere with
accurate gas reading.
In the
Spirometry
menu, if the
Data Source
is set to
Patient
,
no Paw, Agent, or CO2 waveforms or numeric
information is displayed during the airway gas module
warm-up period (approximately 2 minutes). Set the
Data
Source
to
Vent
to display the waveforms and numeric
information from the ventilator during the airway gas
module warm-up period.
Leaks in the gas sampling circuit (water trap and
sampling line) may cause inaccurate readings.
Use only cables and accessories approved by GE
Healthcare Finland Oy. Other cables and accessories
may damage the system or interfere with measurement.
Single-use accessories are not designed to be reused.
Strong scavenging suction on the monitor exhaust port
may change the operation pressure of the monitor and
cause inaccurate readings or internal damage.
Use only airway modules that have anesthetic agent monitoring and
O2 monitoring on this system. Refer to the manual supplied with the
airway modules for detailed information on the modules. The
following modules can be used on this system:
•
E-series: E-CAiO, E-CAiOV, E-CAiOVX (software version 4.5
and above).
•
M-series: M-CAiO, M-CAiOV, M-CAiOVX (software version 3.2
and above).
•
CARESCAPE series: E-sCAiO and E-sCAiOV.
WARNING
•
•
CAUTION
•
Aisys CS²
6-2
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...