4.
To change ventilation mode, select
Mode
. Make the change.To
change the ventilation settings, select a ventilator quick key or
More Settings
. Make the change.
5.
To change alarm settings, select
Alarm Setup
. Make the
change.
6.
To change the gas settings or the circuit type, select
Gas Setup
.
Make the change.
7.
From the
Start Case
menu, select
Start Case Now
. Gas flow
starts.
" section for information on the
Vent Mode
menu.
See the "
" section for information on the
Gas Setup
menu.
" section for information on the
Alarm Setup
menu.
End a case
Use the
End Case
menu to stop gas flow and end the patient
alarms.
1.
Set the Bag/Vent switch to Bag.
2.
Select
End Case
.
3.
Select
End Case Now
on the menu to put the system in standby
(stops the gas flow and patient alarms). The
End Case
menu
shows the gas and agent usage for the case.
Aisys CS²
3-6
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...