Before every patient
This check does not need to be done before the first case of the day
if the "
Every day before your first patient
" checklist was done.
□
Check that the necessary emergency equipment is available
and in good condition.
□
Select
Checkout
and perform a
Low P Leak
test.
□
If the cassette was changed since the last case:
•
Check that the amount of liquid agent in the cassette is
adequate.
•
Check that the cassette is properly inserted. The cassette
is properly inserted when the agent is identified on the
display.
□
Check that an adequate reserve O2 supply is available.
□
Check that the breathing circuit is correctly connected, not
damaged, and the breathing system contains sufficient
absorbent in the canister.
□
Select
Checkout
and perform a
Circuit Leak
test.
□
Check that the ventilator functions correctly:
•
Connect a test lung to the patient breathing circuit
connection.
•
Set the ventilator to VCV mode and the settings for TV to
400 ml, RR to 12, I:E to 1:2, Tpause to Off, and Pmax to
40 cmH2O.
•
Set the gas flow to the minimum settings.
•
Start a case.
•
Set the Bag/Vent switch to Vent.
•
Fill the bellows using O2 flush.
•
Check that mechanical ventilation starts. Check that the
bellows inflate and deflate. Check that the display shows
the correct ventilator data. Check that there are no
inappropriate alarms.
•
Set the appropriate controls and alarm limits for the case.
Make sure that the alarms function. See the "
□
Set the appropriate controls and alarm limits for the case.
Note
4 Preoperative checkout
2067226-001
4-3
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...