1.
Select
System Setup
-
Screen Setup
-
More Settings
or
Spirometry
-
Setup Loops
.
2.
Select
Data Source
.
3.
Select
Patient
to have spirometry data sourced from the airway
module or
Vent
to have spirometry data sourced from the
ventilator.
4.
Select
Back
to view changes made and access other functions
of the
Spirometry
menu.
Setting spirometry volume type
The volume shown on the spirometry split screen can be set to
minute volume or tidal volume.
1.
Select
Spirometry
-
Setup Loops
.
2.
Select
Show MV or TV
and select MV or TV from the drop-
down list.
Set to
TV
to show TVinsp and TVexp on the spirometry split
screen.
Set to
MV
to show MVexp and TVexp on the spirometry split
screen.
3.
Select
Back
to view changes made and access other functions
of the
Spirometry
menu.
Saving, viewing, and deleting spirometry loops
Spirometry loops can be saved, viewed, and deleted through the
Spirometry
menu.
1.
Select
Spirometry
.
2.
To store a loop to memory, select
Save Loop
. Up to six loops
can be saved.
3.
To view a saved loop, set
Show Ref. Loop
to the time at which
it was saved.
4.
To delete a saved loop, set
Delete Ref. Loop
to the time at
which it was saved.
5.
Select
Close
.
Aisys CS²
3-26
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...