Alarms
Alarms may be high priority, medium priority, or low. When an alarm
occurs during a case, an alarm tone sounds and the alarm message
is displayed in the alarm message field. The system checks for alarm
conditions at 1 second intervals. The alarm tone is from 47 to 78
db(A) depending on the alarm volume setting.
No repair should ever be attempted by anyone not having
experience in the repair of devices of this nature. See the
" section.
If an alarm occurs, safeguard the patient first before
performing troubleshooting or doing repair procedures.
Alarm priorities
Alarm priority is indicated by the color of the alarm message and the
audio sequence.
•
High-priority alarm messages appear in white text on a red
background.
•
Medium-priority alarm messages appear in black text on a
yellow background.
•
Low-priority alarms appear in black text on a blue background.
Pausing alarms
Selecting
Audio Pause
for an active alarm stops the audible tone for
120 seconds. The alarm message shows in the alarm message field.
Selecting
Audio Pause
when no medium or high priority alarms are
active prevents the audible alarm tones (audio off) for 90 seconds.
Alarms in the apnea alarm family have special silence behavior to
reduce apnea nuisance alarms. Apnea family alarms include
‘Apnea’, ‘EtCO2 low’, ‘MVexp low’, ‘RR low’, and ‘TVexp low’.
When pausing an apnea family alarm, the audio tone for the active
alarm is paused for 120 seconds. The audible tone for any additional
apnea family alarm that occurs during the audio paused period is
silenced for the remaining time shown on the audio pause
countdown. Only the audible alarm tone is silenced. The alarm
messages still show in the alarm message fields. ‘APN’ shows above
the audio pause countdown when the audible tone silence is in effect
for the apnea family alarms.
CAUTION
WARNING
Aisys CS²
7-2
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...