4.
Texp
5.
PEEP
6.
Flow waveform
7.
TV
Figure 11-3 • Volume control waveform
Pressure control mode (PCV)
Pressure control ventilation supplies a constant set pressure during
inspiration. The ventilator calculates the inspiratory time from the
frequency and I:E ratio settings. A high initial flow pressurizes the
circuit to the set inspiratory pressure. The flow then decreases to
maintain the set pressure (Pinsp).
Pressure sensors in the ventilator measure patient airway pressure.
The ventilator automatically adjusts the flow to maintain the set
inspiratory pressure.
PCV mode settings:
•
Pinsp
•
RR
•
I:E
•
PEEP
•
Pmax
•
Rise Rate
AB.98.037
1
2
3
4
6
5
1.
Paw waveform
2.
Tinsp
3.
Texp
4.
Pinsp
5.
PEEP
Aisys CS²
11-26
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...