Filling Aladin cassettes
Remove the cassette from the machine before filling the cassette.
Make sure that the cassette remains in a horizontal position during
filling. If the cassette is tilted, the flow of liquid is shut off to prevent
overfilling.
Make sure that the valve pins on the back of the cassette do not
come in contact with the system or any other object that could press
on the valve pins and let gas out of the cassette. Observe the liquid
level indicator of the cassette during filling. When the liquid level
reaches the full mark, stop filling.
To avoid spray from anesthetic agent after filling, slowly
remove the agent bottle from the filling port.
Always remove the cassette from the machine and put it
on a horizontal surface before filling. Never try to fill a
cassette while it is in the machine.
Inaccurate anesthetic agent dosages may temporarily
occur after filling if the temperature of the liquid is
considerably different from the normal operating
temperature.
Filling with keyed filler system
AB.60.019
Figure 13-10 • Filling Aladin cassette with keyed filler system
Do not open or press the cassette filling port of the gas
connection valves with fingers or any kind of instrument.
Anesthetic agent liquid or gas may squirt into the air.
1.
Remove the cassette from the machine, and place it on a
horizontal surface. Make sure that the valve pins on the back of
CAUTION
WARNING
•
WARNING
Aisys CS²
13-16
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...