Filling the Reusable Multi Absorber canister
1.
Turn the canister upside down and, using your thumbs, turn the
cover locking ring counterclockwise to unlock it.
AB.74p044
2.
Push up to release the seal.
3.
Lift off the cover to remove it.
AB.74p046
4.
Remove and properly discard the foam filters, the absorbent,
and any water in the reservoir.
Be careful when draining condensate from the
absorber. The liquid is caustic and may burn skin.
--
WARNING
Aisys CS²
8-8
2067226-001