Connecting the airway module
1.
Check that the airway gas module is installed.
2.
Check that the airway adapter connections are tight and that the
adapter is correctly installed.
3.
Check that the water trap container is empty and properly
attached.
Empty the water trap container whenever the container is more
than half full. Under normal conditions, the D-fend container fills
in 24 hours.
Before connecting the exhaust line to the sample gas
outlet on the compact airway module, ensure the other
end is connected to the sample gas return port on the
anesthesia machine. Incorrect connections may cause
patient injury.
--
4.
Connect a line from the airway module sample gas outlet to the
sample gas return port on the anesthesia machine.
5.
Attach the gas sampling line to the sampling line connector on
the water trap.
6.
Turn the system on. The system does a series of automated self
tests. The automatic agent identification is activated.
7.
Connect the sampling line to the airway adapter. Take the gas
sample as close to the patient’s airway as possible. Position the
adapter’s sampling port upwards to prevent condensed water
from entering the sampling line.
1
2
3
AC.22.010
1.
Airway module
2.
Gas sampling line
3.
Airway adapter with sampling line connector
Figure 6-3 • Airway gases setup with compact airway module
WARNING
6 Airway modules
2067226-001
6-5
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...