Circuit compliance compensation
The ventilator adjusts gas delivery and monitoring to compensate for
the compliance of the patient circuit if:
•
The system has an airway module installed.
•
The
Circuit Compliance
is set to
On
from
Super User
-
System Config.
-
Ventilator Settings
.
•
The
Checkout
is completed after the system is turned on.
Circuit compliance is determined during the
Checkout
-
Full Test
or
when the
Vent and Gas
check is done as an individual test. Circuit
compliance must be set by the Super User for the Circuit
Compliance compensation function to be active. See the “Super user
mode” section for information on the circuit compliance setting.
In volume modes, circuit compliance compensation increases the
volume delivered at the inspiratory port by taking into consideration
the circuit compliance value. In all modes, circuit compliance
compensation adjusts the volume measurements. Circuit compliance
compensation provides consistent ventilator accuracy at the patient
circuit.
Perform a
Vent and Gas
test after changing the patient
tube type. Changing the patient breathing circuit after
completing a
Vent and Gas
test affects the volume
measurements in all modes.
Breathing circuits and breathing circuit components are
available in many different configurations from multiple
suppliers. Attributes of the breathing circuits such as
materials, tube length, tube diameter, and configuration
of components within the breathing circuit, may result in
hazards to the patient from increased leakage, added
resistance, or changed circuit compliance.
WARNING
•
Aisys CS²
5-4
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...