Cancelling audio pause
Selecting and holding
Audio Pause
for 2 seconds will cancel the
audio pause function.
Display changes during alarms
Messages may appear in the waveform field during some alarms. If
more than one alarm has a message, the message for the highest
priority alarm is displayed. The message is removed when the alarm
is resolved.
The color of the alarm text shown in the alarm message fields is
dependent on the alarm priority. Messages for high-priority alarms
use white text. Messages for medium-priority alarms use black text.
Low-priority alarm messages use black text.
When the O2 pipeline supply pressure drops to less than 252 kPa
(36 psi), O2 supply information is displayed.
Some patient parameter alarms such as ‘Ppeak high’ and ‘FiO2 low’
will latch when the alarm condition is corrected. When an alarm is
latched, it is displayed in gray text on a black background. The
parameter box will stop flashing. All the associated waveform,
numeric, and digit field messages are removed from the display. The
alarm will remain in this condition until it is acknowledged by
selecting
Audio Pause
or until the alarm reoccurs. When the alarm
is acknowledged, it is removed from the screen. If an alarm has
latched and the alarm reoccurs before it is acknowledged, the alarm
will revert to an active state.
De-escalating alarms
Some device related alarms, such as ‘No insp flow sensor’ will de-
escalate priority when the alarm is acknowledged by selecting Audio
Pause. The alarm message shows at the low-priority alarm level until
the alarm condition is resolved and the alarm is cleared. If that alarm
reoccurs after it has been resolved, the alarm occurs at its standard
priority level.
Battery indicator
The color and fill amount of the battery in use symbol indicates the
amount of battery power remaining.
•
Green indicates greater than 10 minutes.
•
Yellow indicates between 10 and 5 minutes.
7 Alarms and troubleshooting
2067226-001
7-3
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...