2.
Select
Volume Apnea Setup
.
3.
Set
Volume Apnea Selection
to
Enable
or
Disable
.
Set to
Enable
to allow the user to turn the volume apnea alarm
on or off from the
Start Case
menu.
Set to
Disable
to set the volume apnea alarms as always on
during manual ventilation.
4.
Set the Volume Apnea for each case default.
5.
Select
Back
to return to the
Case Defaults
menu.
When the
Volume Apnea Selection
is set to
Disable
, the
Volume
Apnea Alarm
menu selection does not show on the user’s
Start
Case
menu and the
Vol Apnea Alarm
selection does not show on
the
Alarm Setup
menu.
--
Default case type settings
Each case has multiple settings. The default settings for the default
case types are shown in the following table. An * indicates that the
setting is not used for the default ventilation mode in the case type.
Use the empty columns to write in facility changes.
VCV ADULT settings are used as the default if the system does not
have an optional ventilation mode shown.
.
Page 1 Default settings for default case types
Settings
ADULT
PEDIATRIC
LOCAL
CUSTOM 1
Name
ADULT
PEDIATRIC
LOCAL
CUSTOM 1
Patient and Sensor Adult
Pedi
Adult
Adult
Ideal Weight
70
18
70
70
Age
40
5
40
40
Other Gas
Air
Air
Air
Air
Circuit
Circle
Circle
Circle
Circle
Data Source
Vent
Vent
Vent
Vent
O2%
100
100
100
100
Total Flow
6
6
0.2
6
Auto MV Limits
Off
Off
Off
Off
Alarm Volume
3
3
1
3
MV/TV Alarms
On
On
On
On
Note
Note
12 Super user mode
2067226-001
12-11
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...