Inspiratory Breathing resistance in bag mode*
l/min
kPa
cmH2O
5
0.06
0.6
30
0.21
2.1
60
0.49
5.0
*Values include patient circuit tubing and wye-piece 0.049 kPa (0.5
cmH2O) at 60 l/min. Patient circuit tubing and breathing system
configurations affect resistance.
Expiratory Breathing resistance in bag mode*
l/min
kPa
cmH2O
5
0.05
0.5
30
0.15
1.5
60
0.37
3.8
EZchange canister system and condenser, absorber mode
5
0.06
0.6
30
0.19
1.9
60
0.44
4.5
EZchange canister system and condenser, canister removed
5
0.06
0.6
30
0.19
1.9
60
0.44
4.5
*Values include patient circuit tubing and wye-piece 0.03 kPa (0.3 cmH2O)
at 60 l/min. Patient circuit tubing and breathing system configurations
affect resistance.
Pressure flow data (APL valve completely open)
Flow (l/min)
Flow (l/s)
APL pressure cmH2O
3
0.05
0.93
30
0.51
1.70
60
1.0
2.44
70
1.17
3.30
Gas scavenging
All scavenging
Positive pressure relief
10 cmH2O
Aisys CS²
11-14
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...