Setting principal setting defaults
1.
From Super User mode, select
System Config.
and then
Parameter Settings
.
2.
Select the item to change.
For
Principal Volume
, select
MV
or
TV
from the drop-down list.
For
Principal O2
, select
Et
or
Fi
from the drop-down list.
For
Principal Agent
, select
Et
or
Fi
from the drop-down list.
3.
Select
Back
to return to the
System Config.
menu.
Setting volume conditions and humidity
compensation
Individual volume calculation conditions are set for the ventilator and
airway module.
When set to
ATPD
, the volume calculation conditions are based on
ambient temperature and pressure, dry humidity conditions. When
set to
BTPS
, the volume calculation conditions are based on body
temperature, ambient pressure, saturated humidity conditions.
1.
From Super User mode, select
System Config.
and then
Parameter Settings
.
2.
Select the item to change.
•
To set the volume calculation conditions, set
Vent and
Airway Gas Module TV Basis
. Set to
ATPD
/
ATPD
,
ATPD
/
BTPS
, or
BTPS
/
BTPS
.
•
To set the humidity compensation type, set
CO2 Numbers
to
Dry
or
Wet
.
3.
Select
Back
to return to the
System Config.
menu.
Setting inspiratory flow
1.
From Super User mode, select
System Config.
and then
Parameter Settings
.
2.
Select
Insp Flow
to set the graphical display for flow data.
Select the plus symbol to graphically display inspiratory flow as
a positive value.
Select the minus symbol to graphically display the expiratory
flow as a positive value.
3.
Select
Back
to return to the
System Config.
menu.
Note
12 Super user mode
2067226-001
12-7
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...