Alarm setup
Use the
Alarm Setup
menu to set and adjust alarm limits, alarm
volume, and other alarm settings and to view alarm history. The
Alarm Setup
menu contains the
Primary Limits
,
More Limits
,
Alarm History
, and
Configure
submenus.
Setting
Leak Audio
to
Off
silences audio alarms for small leaks.
Leak Audio
is automatically set to
On
and cannot be changed when
either the Low MV alarm limits are off or the
MV/TV Alarms
is set to
Off
.
Selecting
Change to Default Limits
loads the default settings as set
by the Super User or the factory defaults if no Super User settings
have been entered.
Setting CO2 alarms
Use the CO2 Alarms setting to turn off the CO2 ‘Apnea’ alarm,
‘EtCO2 low’, ‘EtCO2 high’, and ‘FiCO2 high. Absorbent OK?’ alarms
during manual ventilation.
CO2 information is obtained from the airway module installed in the
anesthesia system module bay. The CO2 alarms setting has no
effect if there is no airway module in the anesthesia system.
1.
Select
Alarm Setup
.
2.
To turn off the CO2 alarms, set
CO2 Alarms
to
Off
.
The ‘CO2 Alarms Off’ message shows in the general message
field.
The CO2 alarm limit waveform numerics shows dashes during a
case.
The alarms remain disabled until the Bag/Vent switch is set to
Vent, the case is ended, or the
CO2 Alarms
is set to
On
.
3.
To turn on the CO2 alarms, set
CO2 Alarms
to
On
.
4.
Select
Close
.
Setting volume apnea alarm
Use the
Vol Apnea Alarm
setting to turn off the volume apnea alarm
during manual ventilation. The volume apnea alarm remains off until
the Bag/Vent switch is set to Vent or
Vol Apnea Alarm
is set to
On
.
Vol Apnea Alarm
does not show on the
Alarm Setup
menu when
the
Volume Apnea Selection
has been disabled by the Super User.
" section for more information.
1.
Select
Alarm Setup
.
Note
Note
3 Operation
2067226-001
3-17
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...