Electrical connections
Mains inlet
The arrow shows the mains power inlet and cord.
AB.75p008
Equipotential stud
The equipotential stud is used to connect the anesthesia machine to
an equipotential grounding system by attaching a potential
equalization conductor. Equipotential grounding is used in some
hospitals to enhance electrical safety in critical care areas by
attempting to keep the conductive surfaces of all equipment in the
patient care area at the same ground potential, thereby minimizing
unwanted current flow.
Outlets
Labels show outlet voltage ratings and circuit breaker amp ratings.
These are isolated outlets. Regularly test the leakage current.
Aisys CS²
8-10
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...