Manual ventilation cardiac bypass and VCV cardiac
bypass modes should only be used when the patient is
receiving extra-corporeal oxygenation by means of a
heart-lung machine. These modes of ventilation are not
intended to provide metabolic levels of ventilation to the
patient.
Using manual ventilation cardiac bypass
1.
Set the Bag/Vent switch to Bag.
2.
Select
Procedures
.
3.
Select
Start Cardiac Bypass
.
The ‘Cardiac Bypass’ message shows in the waveforms and in
the general message field when manual ventilation cardiac
bypass is active.
4.
Select
Close
.
Using VCV cardiac bypass
1.
Start mechanical ventilation in VCV mode.
2.
Select
Procedures
.
3.
Select
Start Cardiac Bypass
.
PEEP is set to 5 cmH2O.
TV settings of less than 170 ml prior to starting cardiac bypass
remain at the set TV.
TV settings of more than 170 ml prior to starting cardiac bypass
change to 170 ml.
The ‘VCV Cardiac Bypass’ message shows in the waveforms
and in the general message field when VCV cardiac bypass is
active.
PEEP and TV settings can be changed after entering cardiac bypass
mode.
--
4.
Select
Close
.
Vital capacity
Use the Vital Capacity procedure to deliver a pressure breath for a
set time. The Vital Capacity procedure provides a simple way to
deliver one pressure breath during mechanical ventilation without
making multiple ventilator setting changes. The
PEEP on Exit
setting provides a way to change the ventilation PEEP setting
automatically at the end of the Vital Capacity procedure.
WARNING
Note
Aisys CS²
3-28
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...