Gas usage
Use the
Gas Usage
menu to view the total fresh gas usage since the
last reset, to reset the gas usage to zero, and view optional
ecoFLOW and agent cost information.
Resetting cumulative gas usage
1.
From Super User mode, select
Gas Usage
and then
Cumulative Gas Usage
. Gas and agent usage since the last
reset will show.
2.
Select
Reset Usage
. Notice that the gas and agent usage totals
will reset to zero and that
Last Reset
shows the current date.
3.
Select
Back
to return to the
Gas Usage
menu.
Setting ecoFLOW
Use the
ecoFLOW
menu to set the split screen defaults for
FiO2
Setting
,
FiO2 Line
, and
Agent Cost
. This menu item is only
available on systems with the ecoFLOW option installed.
1.
From Super User mode, select
Gas Usage
and then
ecoFLOW
.
2.
Select the item to change. Make the change.
For
FiO2 Setting
, set the numeric value for FiO2 on the split
screen.
For
FiO2 Line
, select whether to show or hide the FiO2 line on
the split screen.
For
Agent Cost
, select whether to show or hide the agent
delivered costs on the split screen.
3.
Select
Back
to return to the
Gas Usage
menu.
Setting agent costs
Use the
Agent Costs
menu to modify the currency and agent cost
defaults used to show the cost of agent delivered on the split screen.
This menu item is only available on systems with the ecoFLOW
option installed.
1.
From Super User mode, select
Gas Usage
and then
Agent
Costs
.
2.
Select the
Currency
from the drop-down list.
3.
Set the cost for each agent.
12 Super user mode
2067226-001
12-3
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...