Setting fresh gas controls
Use the
Fresh Gas Controls
selection to set the gas control style to
O2% with Total Flow or to individual gas flow.
Selecting
O2%
shows O2% as the first quick key and Total Flow
l/min as the second quick key.
Selecting
Flow
shows balance gas l/min as one of the quick keys
and O2 l/min as the other quick key.
Both
O2%
and
Flow
are available if
User
is selected by the Super
User. If
User
is not selected by the Super User, either
O2%
or
Flow
will be unavailable.
Fresh Gas Controls
cannot be accessed during a case.
1.
Select
System Setup
-
Screen Setup
.
2.
Select
More Settings
.
3.
Select
Fresh Gas Controls
.
4.
Select
O2%
or
Flow
from the drop-down menu.
When set to
O2%
, balance gas adjusts automatically when
either the O2% or the Total Flow is changed using the gas quick
keys.
When set to
Flow
, balance gas and the O2% are controlled
individually using the gas quick keys.
5.
Select
Close
.
Fresh gas usage
Use
Fresh Gas Usage
to view the volume of O2, Air, N2O, and
agents used for the three most recent cases.
•
Data only shows for gases available on the system.
•
Agent data shows the three most recently used agents.
1.
Select
System Setup
-
Fresh Gas Usage
.
2.
Select
Case Start Time
to select the patient case to view.
3.
Select
Close
.
Note
Note
Aisys CS²
3-16
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...