Circuit O2 cell
The
Circuit O2 Cell
test measures the O2 percent.
To run this test, follow the on-screen instructions. Do not select
Done
when 21 is first displayed. Allow the reading to stabilize, then
select
Done
. Calibrate the O2 cell if necessary.
External gas monitor
When
External Gas Monitor
is set to
Yes
by the Super User, the
External Gas Monitor
reminder occurs. This is not a test. This is a
reminder to connect a respiratory gas monitor.
5 Preoperative tests
2067226-001
5-7