Test Results
364
88000 Series
Alarms/Errors Overview
Ethernet - PCS Lanes
Note:
Available with parallel interfaces only: 40/100/400G Gbit/s Ethernet and
OTN with 40/100 GbE Ethernet client. Available under the
PCS
sub tab for
Through Mode test providing alarms/errors for both ports.
Skew Alarm Threshold (bits)
, available with Ethernet test application,
allows setting the threshold value that will be used to declare a skew
alarm.
Default
restore the default skew alarm threshold value.
Alarms
Note:
For
LOA
and
Inv. Mapping
alarms see
LOBL
(Loss of Block Lock)
RX
: In lock mode, 65 invalid sync fields (00 or 11) within a 1024 sync
window are received. A PCS Lane Lock alarm is cleared when
receiving 64 consecutive valid 66b sync field (01 or 10) in a row.
LOAML
(Loss of Alignment Marker Lock)
RX
: In lock mode, four consecutive marker values are received that do
not match the alignment marker that the lane is currently locked to.
The LOAML alarm is cleared when the PCS Lane is declared Lock and
two valid alignment markers 16384 blocks (66b) apart are received.
Rate
Range
Default Value
400G
0 to 10550
4781
100G
0 to 2047
928
40G
0 to 4095
1856
Содержание Power Blazer 88000 Series
Страница 1: ...User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE ...
Страница 12: ......
Страница 16: ......
Страница 22: ......
Страница 28: ......
Страница 38: ......
Страница 58: ...Test Setup Test Applications 46 88000 Series OTN BERT ...
Страница 64: ...Test Setup Test Applications 52 88000 Series SONET SDH BERT For SDH BERT ...
Страница 69: ...Test Setup Test Applications Power Blazer 57 SONET SDH DSn PDH BERT For SDH ...
Страница 550: ......
Страница 670: ......
Страница 674: ......
Страница 686: ......
Страница 707: ...Glossary Power Blazer 695 10G Ethernet Client The following figure presents a typical network application ...
Страница 721: ...Glossary Power Blazer 709 Ethernet Cables Straight Through Cable 1000 Mbit s Crossover Cable 1000 Mbit s ...
Страница 765: ...Glossary Power Blazer 753 SONET DSn SDH PDH The GUI Grid indicates the TUG 2 x and TUG 3 x values ...