Fig 71. Centering gasket on bonnet seal surface
1
Gasket
2
Bonnet seal surface
8.
Thread the bonnet onto the body until the body threads are no longer visible.
If the bonnet does not fully thread onto the body, the gasket is not centered on the bonnet seal surface
.
9.
Stabilize the body with a wrench and tighten the bonnet to 62.2 Nm (550 in.-lb). Test for proper operation.
10.3.3
Cleaning the cell mirror
If contamination makes its way into the cell and accumulates on the internal optics, a
DC spectrum power range
exceeded
fault will result. If mirror contamination is suspected, contact Service before attempting to clean the
mirrors. If advised to do so, use the following procedure.
INVISIBLE LASER RADIATION
The sample cell assembly contains a low-power, 10 mW maximum, CW Class 3b invisible laser with a
wavelength between 750-3000 nm. Never open the sample cell flanges or the optical assembly unless the power
is turned off.
This procedure should be used ONLY when necessary and is not part of routine maintenance. To avoid
compromising the system warranty,
Disregard steps 4 and 16 for analyzers without a sample conditioning system (SCS)
.
Tools and materials
•
Lens cleaning cloth (ColeParmer
®
EW-33677-00 TEXWIPE
®
Alphawipe
®
Low-Particulate Clean Room Wipes or
equivalent)
•
Reagent-grade isopropyl alcohol (ColeParmer
®
EW-88361-80 or equivalent)
•
Small drop dispenser bottle (Nalgene
®
2414 FEP Drop Dispenser Bottle or equivalent)
•
Acetone-impenetrable gloves (North NOR CE412W Nitrile Chemsoft™ CE Clean room Gloves or equivalent)
•
Hemostat (Fisherbrand™ 13-812-24 Rochester-Pean Serrated Forceps)
•
Bulb blower or dry compressed air/nitrogen
•
Torque wrench
•
3 mm hex driver
•
Non-out-gassing grease
•
Flashlight
Process samples may contain hazardous material in potentially flammable and/or toxic concentrations.
Personnel should have a thorough knowledge and understanding of the physical properties and safety
precautions for the sample contents before operating the SCS.
All valves, regulators, switches, etc. should be operated in accordance with site lockout/tagout procedures.
1.
Power down the analyzer.
2.
Isolate the SCS from the process sample tap.
3.
If possible, purge the system with nitrogen for 10 minutes.