Operating Instructions
J22 TDLAS Gas Analyzer
130
Hauser
11.6.12
ATEX/IECEx heater, E+H material number 70188857 (SS P/N 1100002271)
Materials
1.
Socket head screw, M5-0.8 x 50 (3)
2.
Lock washer (3)
3.
Flat washer (3)
4.
Heater
5.
Label, terminal block
6.
Ground cable GRN/YEL
7.
Locking terminal fork (6)
8.
Thermostat
9.
Thermal compound grease
10.
Label, terminal block
11.
Terminal block
Torque screws (item 1) to 5.1 Nm (45.1 lb-in).
Crimp terminals per manufacturer specifications with
Panduit CT-1550 or equivalent.
Apply a thin, even coat of thermal compound (item 9) 0.1 mm thick to the bottom surface of the heater block
(item 4) when installing it on the heater plate assembly
Service technicians will be required to install the input power.
Refer to the dashed lines in the wiring diagram for field technicians installation requirements and the solid lines
for the factory installed components.
Both heater and thermostat ground wires use the same terminal fork.
11.6.13
CSA heater, E+H material number 70188858 (SS P/N 1100002272)
WIRING ASSEMBLY:
(REF WIRING SCHEMATIC BELOW)
GROUND
SCREW
A 100 to 240 VAC ± 10%, 50/60 HZ, Main power
B Heater
G/Y
Green/yellow
C
Thermostat
L
Line
BR Brown wire
N
Neutral
BL Blue wire
G
Ground
A
L BR
N BL
BROWN BL
G G/Y
G/Y
Wiring diagram
B
C
WIRING ASSEMBLY:
(REF WIRING SCHEMATIC BELOW)
GROUND