J22 TDLAS Gas Analyzer
Operating Instructions
Hauser
73
7.15.1
Stream submenu
In the stream submenu, you can set parameters related to the stream that needs to be measured.
Navigation
Setup menu → Advanced setup → Stream
▸
Stream
Analyte type
Select calibration
Rolling average number
Parameter overview with brief description
Parameter
Description
User entry
Factory setting
Analyte type
The analyte measured by the
analyzer.
—
H
2
O
Select calibration
Changes and sets the calibration •
1
•
2
•
3
•
4
1
Rolling average
number
Sets the number of
measurements included in the
running average.
Positive integer
4
7.15.2
Sensor adjustment submenu
The sensor adjustment submenu contains parameters that pertain to the functionality of the sensor.
Navigation
Setup menu → Advanced setup → Sensor adjustment
▸
Sensor adjustment
Concentration adjust
Concentration multiplier (RATA)
Concentration offset (RATA)
2fbase curve source
2fbase curve RT update
▸
Calibration 1 to n
Parameter overview with brief description
Parameter
Prerequisite
Description
User entry
Factory setting
Concentration adjust —
Enables or disables
adjustment factors.
•
On
•
Off
Off
Concentration
multiplier (RATA)
Used if Concentration
Adjust is enabled.
Slope adjustment
factor.
Signed floating point
number
1.0
Concentration offset
(RATA)
Used if Concentration
Adjust is enabled.
Offset adjustment
factor.
Signed floating point
number
0
2fbase curve source
Used if Base Curve
subtraction is enabled.
Selects which reference
to subtract.
•
Ref0Curve
•
Ref0RTCurve
Ref0Curve
2fbase curve RT
update
Used if Base Curve
subtraction is enabled.
Option to update the
stored RT base curve
•
Cancel
•
Start
Cancel