J22 TDLAS Gas Analyzer
Operating Instructions
Hauser
95
Diagnostic
number
Short text
Remedy instructions
Status signal
(from factory)
Diagnostic
behavior
(from factory)
262
Sensor electronic
connection faulty
1.
Check or replace connection cable between
sensor electronic module (ISEM) and main
electronics.
2.
Check or replace ISEM or main electronics.
F
Alarm
270
Main electronic failure
Change main electronic module.
F
Alarm
271
Main electronic failure
1.
Restart device.
2.
Change main electronic module.
F
Alarm
272
Main electronic failure
1.
Restart device.
2.
Contact service.
F
Alarm
273
Main electronic failure
Change electronic.
F
Alarm
275
I/O module 1 to n
defective
Change I/O module.
F
Alarm
276
I/O module 1 to n faulty 1.
Restart device.
2.
Change I/O module.
F
Alarm
283
Memory content
1.
Reset device.
2.
Contact service.
F
Alarm
300
Sensor electronics
(ISEM) faulty
1.
Restart device.
2.
Replace sensor electronics.
F
Alarm
301
SD memory card error
1.
Check SD card.
2.
Restart device.
C
Warning
302
Device verification in
progress
Device verification active, please wait.
C
Warning
303
I/O @1 configuration
changed
1.
Apply I/O module configuration (parameter
'
Apply I/O configuration
').
2.
Afterwards reload device
3.
description and check wiring.
M
Warning
311
Electronic failure
1.
Do not reset device.
2.
Contact service.
M
Warning
330
Flash file invalid
1.
Update firmware of device.
2.
Restart device.
M
Warning
331
Firmware update failed
1.
Update firmware of device.
2.
Restart device.
F
Warning
332
Writing in HistoROM
backup failed
Replace User interface board Ex d/XP: replace
controller
F
Alarm
361
I/O module 1 to n faulty 1.
Restart device.
2.
Check electronic modules.
3.
Change I/O Module or main electronics.
F
Alarm
372
Sensor electronics
(ISEM) faulty
1.
Restart device.
2.
Check if failure recurs.
3.
Replace ISEM.
F
Alarm
373
Sensor electronic (ISEM)
faulty
1.
Transfer data or reset device.
2.
Contact service.
F
Alarm
375
I/O - 1 to n
communication failed
1.
Restart device.
2.
Check if failure recurs.
3.
Replace module rack inclusive electronic
modules.
F
Alarm
382
Data storage
1.
Insert T-DAT.
2.
Replace T-DAT.
F
Alarm
383
Memory content
1.
Restart device.
2.
Delete T-DAT from the '
Reset device
' parameter.
F
Alarm