Operating Instructions
J22 TDLAS Gas Analyzer
86
Hauser
Parameter overview with brief description
Parameter
Prerequisite
Description
User interface/User entry
Factory setting
Current input
1 to n
simulation
—
Switch the simulation of
the current output on and
off.
•
Off
•
On
Off
Value current
input 1 to n
In the
Current input 1 to n
simulation
parameter, the
On
option is selected.
Enter the current value
for simulation.
0 to 22.5 mA
Set to the actual
input current when
the simulation is set
to On.
Current output
1 to n
simulation
—
Switch the simulation of
the current output on and
off.
•
Off
•
On
Off
Current output
value 1 to n
In the
Current output 1 to
n simulation
parameter,
the
On
option is selected.
Enter the current value
for simulation.
3.59 to 22.5 mA
3.59 mA
Switch output
simulation 1 to
n
In the
Operating mode
parameter, the
Switch
option is selected.
Switch the simulation of
the switch output on and
off.
•
Off
•
On
Off
Switch state 1
to n
—
Select the status of the
status output for the
simulation.
•
Open
•
Closed
Open
Relay output 1
to n simulation
—
—
•
Off
•
On
Off
Switch state 1
to n
In the
Switch output
simulation 1 to n
parameter, the
On
option
is selected.
—
•
Open
•
Closed
Open
Device alarm
simulation
—
Switch the device alarm
on and off.
•
Off
•
On
Off
Diagnostic
even category
—
Select a diagnostic event
category.
•
Sensor
•
Electronics
•
Configuration
•
Process
Process
Diagnostic
event
simulation
—
Select a diagnostic event
to simulate this event.
•
Off
•
Diagnostic event picklist
(depends on the
category selected)
Off
8.5
Protecting settings from unauthorized access
The following write protection options exist in order to protect the J22 TDLAS Gas Analyzer software configuration
from unintentional modification:
•
Protect access to parameters with access code
•
Protect access to local operation with
•
Protect access to measuring device with
8.5.1
Write protection through access code
By enabling user-specific access code, the parameters for the measuring device configuration are write-protected and
their values can no longer be changed through local operation.
8.5.2
Defining the access code through local display
1.
Navigate to the
Define access code
parameter.
2.
Define a max. 16-digit character string comprising numbers, letters and special characters as the access code.