J22 TDLAS Gas Analyzer
Operating Instructions
Hauser
61
Setup
Relay output 1 to n
Display
Advanced setup
7.3
Defining the tag name
To enable fast identification of the measuring point within the system, you can enter a unique designation using the
Device tag parameter and thus change the Factory setting.
A0029422
Fig 58. Header of the operational display with tag name
1
Tag name
Navigation
Setup menu → Device tag
Parameter overview with brief description
Parameter
Description
User entry
Factory setting
Device tag
Enter the name for the measuring point.
Max. 32 characters, such as number or special
characters (e.g., @, %, /)
H
2
O analyzer
7.4
Setting the analyte type
Sets the analyte type measured by the analyzer.
Navigation
Setup menu → Analyte type
Parameter overview with brief description
Parameter
Description
User entry
Factory setting
Analyte type
The analyte measured by the analyzer.
—
H
2
O
7.5
Selecting the measurement calibration
Select the calibration to be measured for your device.
Navigation
Setup menu → Select calibration
Parameter overview with brief description
Parameter
Description
User entry
Factory setting
Select calibration
Select the calibration for measurement. (User defined.) In
most cases, calibrations are as follows:
1) Process stream as defined by customer order
2) Methane background
3) Nitrogen background
4) Not used
•
1
•
2
•
3
•
4
1