J22 TDLAS Gas Analyzer
Operating Instructions
Hauser
69
Parameter
Prerequisite
Description
User entry
Factory setting
•
Dew point 2
1
Assign status The
Status
option is selected
in the
Switch output function
parameter.
Select the device status for switch
output.
•
Off
•
Validation
control
Off
Switch-on
value
In the
Switch output
function
parameter, the
Limit
option is selected.
Enter measured value for the
switch-on point.
Signed floating-
point number
0 ppmv
Switch-off
value
In the
Switch output
function
parameter, the
Limit
option is selected.
Enter measured value for the
switch-off point.
Signed floating-
point number
0 ppmv
Switch-on
delay
The
Limit
option is selected
in the
Switch output function
parameter.
Define delay for the switch-on of
status output.
0.0 to 100.0 s
0.0 s
Switch-off
delay
The
Limit
option is selected
in the
Switch output function
parameter.
Define delay for the switch-off of
status output.
0.0 to 100.0 s
0.0 s
Invert output
signal
—
Invert the output
signal.
•
No
•
Yes
No
7.13
Configuring the relay output
The Relay output wizard guides the user systematically through all the parameters that have to be set for configuring
the relay output.
Navigation
Setup menu → Relay output 1 to n
▸
Relay output 1 to n
Relay output function
Terminal number
Assign limit
Assign diagnostic behavior
Assign status
Switch-off value
Switch-on value
→
Switch-off delay
→
Switch-on delay
→
Failure mode
→
Parameter overview with brief description
Parameter
Prerequisite
Description
User entry
Factory setting
Relay output
function
—
Select the function for the
relay output.
•
Closed
•
Open
•
Diagnostic behavior
•
Limit
•
Status
Diagnostic
behavior