Hauser
4.
Plug in the RJ45 connector.
5.
Firmly tighten the cable glands.
This concludes the Modbus TCP connection process.
A033722
Fig 25. Connecting the RJ45 cable
6.
Close the terminal cover.
7.
Fit the display module holder in the electronics compartment.
8.
Screw on the connection compartment cover.
9.
Secure the securing clamp of the connection compartment cover.
4.6.4
Connecting the supply voltage and additional inputs/outputs
The temperature of the J22 TDLAS Gas Analyzer can reach 67
°
C in 60
°
C ambient at the cable entry and
branching point.
These temperatures must be considered when selecting field wiring and cable entry devices.
The main electronic assembly shall be protected by a building installation overcurrent protection rated for 10
amp or less.
1.
Push the cable through the cable entry. To ensure tight sealing, do not remove the sealing ring from the cable
entry.
2.
Strip the cable and cable ends. In the case of stranded cables, also fit ferrules.
3.
Connect the protective ground.
A029815
Fig 26. Feeding wiring and connecting the protective ground
5
1
2
4
3