Operating Instructions
J22 TDLAS Gas Analyzer
82
Hauser
Navigation
Diagnostics menu → Data logging
▸
Data logging
Assign channel 1 to n
Logging interval
Clear logging data
Data logging
Logging delay
Data logging control
Data logging status
Entire logging duration
Parameter overview with brief description
Parameter
Prerequisite
Description
User interface/User
entry
Factory
setting
Assign
channel 1
to n
The
Extended HistoROM
application package is
available.
Assign process variable to
logging channel.
•
Off
•
Concentration
•
Dew point 1
•
Dew point 2
•
Cell gas
pressure
•
Cell gas
temperature
•
Flow switch state
•
Current output 1 to
n
Off
Logging
interval
The
Extended HistoROM
application package is available.
Define the logging interval for
data logging. This value defines
the time interval between the
individual data points in the
memory.
0.1 to 999.0 s
1.0 s
Clear logging
data
The
Extended
HistoROM
application package is
available.
Clear the entire logging data.
•
Cancel
•
Clear data
Cancel
Data logging —
Select the data logging method.
•
Overwriting
•
Not overwriting
Overwriting
Logging
delay
In the
Data logging
parameter,
the
Not overwriting
option is
selected.
Enter the time delay for
measured value logging.
0 to 999 h
0 h
Data logging
control
In the
Data logging
parameter,
the
Not overwriting
option is
selected.
Start and stop measured value
logging.
•
None
•
start
•
Stop
None
Data logging
status
In the
Data logging
parameter,
the
Not overwriting
option is
selected.
Displays the measured value
logging status.
•
Done
•
Delay active
•
Active
•
Stopped
Done
1