Operating Instructions
J22 TDLAS Gas Analyzer
94
Hauser
9.6
Overview of diagnostic information
The amount of diagnostic information and the number of measured variables affected increase if the measuring
device has one or more application packages. In the case of some items of diagnostic information, the Diagnostic
behavior can be changed. See
Adapting the diagnostic information
Diagnostic
number
Short text
Remedy instructions
Status signal
(from factory)
Diagnostic
behavior
(from factory)
Diagnostic of sensor
082
Data storage
1.
Check module connections.
2.
Contact service.
F
Alarm
083
Memory content
1.
Restart device.
2.
Restore HistoROM S-DAT backup. ('
Device reset
'
parameter)
3.
Replace HistoROM S-DAT.
F
Alarm
100
Laser off
1.
Restart device.
2.
Replace Sensor electronics.
3.
Replace sensor (OH).
F
Alarm
101
Laser off
1.
Wait for laser to heat up to temperature.
2.
Replace sensor (OH).
F
Alarm
102
Laser temperature
sensor faulty
1.
Restart device.
2.
Replace Sensor electronics.
3.
Replace sensor (OH).
C
Warning
103
Laser temperature
unstable
1.
Check ambient temperature ramp meets
specification.
2.
Replace sensor electronics.
3.
Replace sensor (OH).
F
Alarm
104
Laser temperature
settling
Wait for laser temperature to settle.
C
Warning
105
Cell pressure connection
defective
1.
Check connection to pressure cell.
2.
Replace pressure cell.
F
Alarm
106
Sensor (Optical Head)
faulty
1.
Restart device.
2.
Replace sensor (OH).
F
Alarm
107
Detector zero range
exceeded
1.
Check process.
2.
Check spectrum.
M, C
Warning
108
Detector reference level
range exceeded
1.
Check process.
2.
Check spectrum.
M, C
Warning
109
Peak index @1 out of
range
1.
Check process.
2.
Check spectrum.
3.
Reset peak tracking.
F
Alarm
110
Peak track adjustment
exceeded
1.
Check process.
2.
Check spectrum.
3.
Reset peak tracking.
F
Alarm
111
Peak track adjustment
warning
1.
Check process.
2.
Check spectrum.
3.
Reset peak tracking.
F
Alarm
Diagnostic of electronics
201
Device failure
1.
Restart device.
2.
Contact service.
F
Alarm
242
Software incompatible
1.
Check software.
2.
Flash or change main electronics module.
F
Alarm
252
Modules incompatible
1.
Check electronic modules.
2.
Change electronic modules.
F
Alarm