Operating Instructions
J22 TDLAS Gas Analyzer
118
Hauser
11.5.4
0.8m cell tube and mirror, E+H material number 70188821 (SS P/N 1100002235)
Materials
1.
Cell tube assembly, 0.8 m
2.
Socket head screw, M4-0.7 x 16 (4)
3.
Lock washer (4)
4.
O-ring, Viton
5.
Tapered vinyl plug
6.
Vinyl cap
When installing the cell tube assembly onto the analyzer, torque the screws (item 2) to 4.5 Nm (39.8 lb-in).
Grease the O-ring (item 4) with Syntheso Glep 1 or equivalent prior to installing.
NACE and MTR reports available upon request.
11.5.5
Flat mirror, E+H material number 70188822 (SS P/N 1100002236)
Materials
1.
Mirror, 0.8 m
2.
Socket head screw, M4-0.7 x 14(3)
3.
Lock washer (3)
4.
O-ring, Viton
When installing the mirror onto the cell tube assembly, torque the screws (item 2) to 2.6 Nm (23 lb-in).
Grease the O-ring (item 4) with Syntheso Glep 1, or equivalent, prior to installing.
NACE and MTR reports available upon request.
11.5.6
Calibrated optical head, E+H material number 70188824 (SS P/N 1100002238)