11.6
Sample conditioning system spare parts details
11.6.1
Analyzer gas fittings, E+H material number 1100002256 (SS P/N 1100002256)
A, A-1
Detail of area A
B, B-1
Detail of area B
C, C-1
Detail of area C
D, D-1
Detail of area D
Materials
1.
Connector fitting
2.
Sealing washer
3.
Hollow hex plug. 1/8 in. NPTM.
Item 3 is located behind 1 and 2 in A-1 on cell tube.
4.
Sealing hex plug M12 x 1.5, O-ring (3)
5.
¼ TF plug (2)
6.
Connector fitting
7.
Tape, third seal TFE
8.
Connector fitting
9.
1/8 in. plug
Use 2 to 3 wraps of tape (item 8) on all connectors and plugs during installation.
Torque hollow hex plug (item 3) to 2.6 Nm (23 lb-in).
Torque sealing hex plug to 7.0 Nm (62 lb-in).
NACE and MTR reports available upon request.
CRN compliant component.
11.6.2
¼ in. I/O connector with purge, E+H material number 1100002257 (SS P/N
1100002257)
Materials
1.
Tube, union bulkhead ¼ TF (6)
Torque nut on ¼ in. swaged bulkhead to 12.0 Nm (106 lb-in.).
NACE and MTR reports available upon request.
CRN compliant component.