J22 TDLAS Gas Analyzer
Operating Instructions
Hauser
133
11.6.17
Krohne armored flow meter, SS P/N 1100002279
A
Hardware for bracket to panel
Materials
1.
Flow meter, armored
2.
Flat washer (2)
3.
Lock washer (2)
4.
Socket head screw, M4-0.7 x 10 (2)
Torque screws (item 4) to 2.6 Nm (23 lb-in).
CRN compliant component.
11.6.18
ATEX Krohne armored flow meter kit, SS P/N 1100002281
A
Both the blue and white wires will have 2 in. of heat-shrink (item 7) installed over the
ends.
B
Hardware for bracket to panel.
BR Brown wire to pin 2 on rectangular connector.
R
Red wire to pin 2 on rectangular connector.
Materials
1.
Flow meter, armored, ATEX
2.
Flat washer (2)
3.
Lock washer (2)
4.
Socket head screw, M4-0.7 x 10 (2)
5.
Contact connector
6.
Cable gland
7.
Heat-shrink, olefin
8.
Rectangular connector, 4 position
Torque screws (item 4) to 2.6 Nm (23 lb-in).
Flowrate range: 0.2-2.000 slpm
J6 Rectangular connector is inserted to the second connector
on the optical head PCB assembly.
Содержание SpectraSensors J22 TDLAS
Страница 2: ...Operating Instructions J22 TDLAS Gas Analyzer 2 Endress Hauser...