Operating Instructions
J22 TDLAS Gas Analyzer
134
Hauser
11.6.19
CSA Krohne armored flow meter kit, SS P/N 1100002282
A
Both the blue and white wires will have 2 in. of heat-shrink (item 7) installed over the ends.
B
Hardware for bracket to panel.
BR Brown wire to pin 2 on rectangular connector.
R
Red wire to pin 2 on rectangular connector.
Materials
1.
Flow meter, armored, CSA
2.
Flat washer (2)
3.
Lock washer (2)
4.
Socket head screw, M4-0.7 x 10 (2)
5.
Contact connector
6.
Cable gland
7.
Heat-shrink, olefin
8.
Rectangular connector, 4 position
Torque screws (item 4) to 2.6 Nm (23 lb-in).
Flowrate range: 0.2-2.000 slpm
CRN compliant component.
11.6.20
Flow meter gas fittings without bypass, SS P/N 1100002283
Materials
1.
Male elbow
2.
Connector fitting
3.
Tape, third seal TFE
Select this connector kit if the sample conditioning system has one flow meter (without bypass).
Apply 2 to 3 wraps of tape (item 3) to both connectors during installation.
NACE and MTR reports available upon request.
CRN compliant component.
J6 Rectangular connector is inserted to the second connector
on the optical head PCB assembly.