J22 TDLAS Gas Analyzer
Operating Instructions
Hauser
131
Materials
1.
Socket head screw, M5-0.8 x 50 (3)
2.
Lock washer (3)
3.
Flat washer (3)
4.
Heater
5.
Terminal block label
6.
Ground cable
7.
Locking terminal fork (6)
8.
Thermostat
9.
Thermal compound grease
10.
Terminal block label
11.
Terminal block
Torque screws (item 1) to 5.1 Nm (45.1 lb-in).
Crimp terminals per manufacturer specifications with
Panduit CT-1550 or equivalent.
Apply a thin, even coat of thermal compound (item
12) 0.1 mm thick to the bottom surface of the heater
block (item 4) when installing it on the heater plate assembly
Service technicians will be required to install the input power.
Refer to the dashed lines in the wiring diagram for field technicians installation requirements and the solid lines
for the factory installed components.
Both heater and thermostat ground wires use the same terminal fork.
11.6.14
King glass flow meter, SS P/N 1100002276
A
Hardware is used to mount the flow meter to the bracket and the bracket to the panel.
Materials
1.
Flow meter, King, glass
2.
Flat washer (4)
3.
Lock washer (4)
4.
Socket head screw, M4-0.7 x 10 (4)
Torque screws (item 4) to 2.6 Nm (23 lb-in).
A 100 to 240 VAC ± 10%, 50/60 HZ, Main power
B Heater
G/Y
Green/yellow
C
Thermostat
L
Line
BR Brown wire
N
Neutral
BL Blue wire
G
Ground
A
L BR
N BL
BR
BL
G G/Y
Wiring diagram
B
C