Operating Instructions
J22 TDLAS Gas Analyzer
66
Hauser
Parameter overview with brief description
Parameter
Prerequisite
Description
User entry
Factory setting
Current span —
Select current range for
process value output and
upper/lower level for alarm
signal.
•
4...20 mA
•
4...20 mA NE
•
4...20 mA US
•
0...20 mA
Approval-specific:
•
4...20 mA NE
•
4...20 mA US
Terminal
number
—
Shows the terminal
numbers used by the
current input module.
•
Not used
•
24-25 (I/O 2)
•
22-23 (I/O 3)
—
Signal mode
The measuring device is
not approved for use in the
hazardous area with Ex-i
protection type.
Select the signal mode for
the current input.
•
Passive
•
Active
Passive
0/4 mA
value
—
Enter 4 mA value.
Signed floating-point
number
Approval-specific:
•
mbar a
•
psi a
20 mA value —
Enter 20 mA value.
Signed floating-point
number
Approval-specific:
•
mbar a
•
psi a
Failure mode —
Define input behavior in
alarm condition.
•
Alarm
•
Last valid value
•
Defined value
Alarm
Failure
current
In the
Failure mode
parameter, the
Defined
value
option is selected.
Enter the value to be used
by the device if input value
from external device is
missing.
Signed floating-point
number
0
7.11
Configuring the current output
The Current output
wizard systematically guides you through all the parameters that have to be set for configuring
the current output.
Navigation
Setup menu → Current output
▸
Current output 1 to n
Pro.var. outp
Terminal number
Current range output
Signal mode
Lower range value output
Upper range value output
Damping current
Fixed current
Fail.behav.out
Failure current
Содержание SpectraSensors J22 TDLAS