Operating Instructions
J22 TDLAS Gas Analyzer
52
Hauser
Measuring device
Settings
Interface
CDI-RJ45
Measuring device
The measuring device has an RJ45 interface.
Web server
Web server must be enabled; Factory setting: ON. For information on
IP address
If the IP address of the device is not known:
•
The IP address can be read out from local operation:
Diagnostics → Device information → IP
address
•
Communication with the web server can be established from the default IP address
192.168.1.212. The DHCP function is enabled in the device at the factory, i.e. the device
expects an IP address to be assigned by the network. This function can be disabled and the
device can be set to the default IP address 192.168.1.212: set DIP switch No. 4 from OFF →
ON.
Set the default IP address →
5.6.2
Connecting to the analyzer through the service interface (CDI-RJ45)
Preparing the measuring device
1.
Loosen the securing clamp of the connection compartment cover.
2.
Unscrew the connection compartment cover.
3.
Unhitch the display module and place on the side of the controller housing, then open the transparent shield
cover from the RJ45 connector.
4.
Connect the computer to the RJ45 connector from the standard Ethernet connecting cable.
A0027563
Fig 52. Connecting through CDI-RJ45
1
Computer with web browser for accessing the integrated device web server
2
Standard Ethernet connecting cable with RJ45 connector
3
Service interface (CDI-RJ45) of the measuring device with access to the integrated web server
Configuring the Internet protocol of the computer
The measuring device works with the Dynamic Host Configuration Protocol (DHCP) on leaving the factory. The IP
address of the measuring device is automatically assigned by the automation system (DHCP server).
The IP address can be assigned to the measuring device in a variety of ways:
•
Dynamic Host Configuration Protocol (DHCP), Factory setting:
The IP address is automatically assigned to the
measuring device by the automation system (DHCP server).
•
•
Software addressing:
The IP address is entered from the
•
DIP switch for Default IP address:
To establish the network connection through the
: the fixed IP address 192.168.1.212 is used.
The following information refers to the default Ethernet settings of the device.