J22 TDLAS Gas Analyzer
Operating Instructions
Hauser
139
Item
Description
Sample process temperature
-20 °C to 60 °C (-4 °F to 140 °F)
Sample flow rate
0.5 to 1.0 slpm (1 to 2 scfh)
Bypass flow rate
0.5 to 1.0 slpm (1 to 2 scfh)
Process seal
Dual Seal without Annunciation
Primary process seal 1
UV Grade Fuse Silica Glass
Primary process seal 2
Primary Process Seal 2
Secondary process seal
Elastosil RT 622
12.3
Physical specificiations
Item
Description
Weight
J22 TDLAS Gas Analyzer
: 16 kg (36 lbs)
J22 TDLAS Gas Analyzer with SCS
on Panel
: 24 kg (53 lbs)
J22 TDLAS Gas Analyzer
with Enclosed SCS
1
: 43 kg (95 lbs)
J22 TDLAS Gas Analyzer with Enclosed SCS
1
, Heated
: 43 kg (95 lbs)
Dimensions
J22 TDLAS Gas Analyzer
CSA: 727 mm H x 236.2 mm D x 224 mm W (28.6 in. H x 9.3 in. D x 8.8 in. W)
ATEX: 727 mm H x 236.2 mm D x 192 mm W (28.6 in. H x 9.3 in. D x 7.5 in. W)
J22 TDLAS Gas Analyzer with SCS
1
on Panel
737 mm H x
241 mm D x 376 mm W (29 in. H x 9.5 in. D x 14.8 in. W)
J22 TDLAS Gas Analyzer
with Enclosed SCS
1
/
J22 TDLAS Gas Analyzer with Enclosed SCS
1
, Heated
838 mm H x
255 mm D x 406 mm W (33 in. H x 10 in. D x 16 in. W)
12.4
Area classification
Item
Description
J22 TDLAS Gas Analyzer
cCSAus: Ex db ia [ia Ga] op is IIC T4 Gb
Class I, Zone 1, AEx db ia [ia Ga] op is IIC T4 Gb
Class I, Division 1, Groups A, B, C, D, T4
Tambient = -20
°
C to 60
°
C
ATEX/IECEx/UKEX:
II 2G Ex db ia [ia Ga] ib op is IIC T4 Gb
Tambient = -20
°
C to 60
°
C
IECEx (PESO): Ex db ib op is IIC T4 Gb
JPN: Ex db ia [ia Ga] ib op is IIC T4 Gb
KTL: Ex db ia [ia Ga] ib op is IIC T4 Gb
INMETRO: Ex db ia [ia Ga] ib op is IIC T4 Gb
Tambient = -20 °C to 60 °C
J22 TDLAS Gas Analyzer with SCS
on panel
cCSAus: Ex db ia op is IIC T4 Gb
Class I, Zone 1, AEx db ia op is IIC T4 Gb
Class I, Division 1, Groups A, B, C, D, T4
1
Sample conditioning system
2
Sample conditioning system