J22 TDLAS Gas Analyzer
Operating Instructions
Hauser
65
Parameter
Prerequisite
Description
User entry
Factory
setting
Parity
Modbus RS485 device
Select parity bits.
Picklist ASCII
option:
•
0 = Even option
•
1 = Odd option
Picklist RTU option:
•
0 = Even option
•
1 = Odd option
•
2 = None / 1
stop bit option
•
3 = None / 2
stop bits option
Even
Byte order
Both Modbus RS485
and Modbus TCP
Select byte transmission sequence.
•
0-1-2-3
•
3-2-1-0
•
1-0-3-2
•
2-3-0-1
1-0-3-2
Prio. IP
address
Modbus TCP device
The IP address for which connections are
accepted by the priority pool.
IP address
0.0.0.0
Inactivity
timeout
Modbus TCP device
Time before a connection can be terminated due
to inactivity. A setting of zero means no timeout.
0 to 99 seconds
0
seconds
Max
connections
Modbus TCP device
Maximum number of concurrent connections.
Priority pool connections take precedence and are
never denied connection thus causing the oldest
connection to be terminated.
1 to 4
4
Failure mode Both Modbus RS485
and Modbus TCP
Select measured value output behavior when a
diagnostic message occurs through Modbus
communication.
Not a Num (NaN).
—
—
7.10
Configuring the current input
The Current input wizard guides the user systematically through all the parameters that have to be set for
configuring the current input.
Navigation
Setup menu → Current input
▸
Current input 1 to n
Current span
Terminal number
Signal mode
0/4 mA value
20 mA value
Failure mode
Failure current