Operating Instructions
J22 TDLAS Gas Analyzer
90
Hauser
9.2
Diagnostic information on local display
9.2.1
Diagnostic message
Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in
alternation with the operational display.
A0029426-EN
Fig 64. Diagnotic message
1
Status signal
2
Diagnostic behavior
3
Diagnostic behavior with diagnostic code
4
Short text
5
Operating elements
If two or more diagnostic events are pending simultaneously, only the message of the diagnostic event with the
highest priority is shown.
Other diagnostic events that have occurred can be displayed in the
Diagnostics
menu:
•
From parameters
•
9.2.1.1
Status signals
The Status signals provide information on the state and reliability of the device by categorizing the cause of the
diagnostic information (diagnostic event). The Status signals are categorized according to
VDI/VDE 2650
and
NAMUR Recommendation NE 107
: F = Failure, C = Function Check, S = Out of Specification, M = Maintenance
Required.
Symbol
Meaning
F
Failure.
A device error has occurred. The measured value is no longer valid.
C
Function check.
The device is in service mode (e.g., during a simulation).
S
Out of specification.
The device is operated outside its technical specification limits (e.g., outside the process
temperature range)
M
Maintenance required.
Maintenance is required. The measured value remains valid.
9.2.1.2
Diagnostic behavior
Symbol
Meaning
Alarm.
Measurement is interrupted. Signal outputs assume the defined alarm condition. A diagnostic message is
generated.
Warning.
Measurement is resumed. The signal outputs are not affected. A diagnostic message is generated.