J22 TDLAS Gas Analyzer
Operating Instructions
Hauser
51
5.6
Access to the operating menu from the web browser
Thanks to the integrated web server, the device can be operated and configured through a web browser through a
service interface (CDI-RJ45) and connected for Modbus TCP signal transmission. The structure of the operating menu
is the same as for the local display. In addition to the measured values, status information on the device is displayed
and allows the user to monitor the status of the device. The measuring device data can also be managed and the
network parameters can be configured.
5.6.1
Prerequisites
Computer hardware
Hardware
Interface
CDI-RJ45
Interface
The computer must have an RJ45 interface.
Connection
Standard Ethernet cable with RJ45 connector.
Screen
Recommended size: ≥12 in. (depends on the screen resolution)
Computer software
Software
Interface
CDI-RJ45
Recommended operating
systems
•
Microsoft Windows 7 or higher.
•
Mobile operating systems:
•
iOS
•
Android
Web browsers supported
•
Microsoft Internet Explorer 8 or higher
•
Microsoft Edge
•
Mozilla Firefox
•
Google Chrome
•
Safari
Computer settings
Settings
Interface
CDI-RJ45
User rights
Appropriate user rights (e.g., administrator rights) for TCP/IP and proxy server
settings are necessary (for adjusting the IP address, subnet mask, etc.).
Proxy server settings of the
web browser
The web browser setting
Use a Proxy Server for Your LAN
must be
deselected.
JavaScript
JavaScript must be enabled.
If JavaScript cannot be enabled, enter http://192.168.1.212/basic.html in the address line
of the web browser. A fully functional but simplified version of the operating menu
structure starts in the web browser.
When installing a new firmware version: To enable correct data display, clear the
temporary memory (cache) of the web browser under
Internet options
.
Network connections
Only the active network connections to the measuring device should be used.
Switch off all other network connections such
as WLAN.
Switch off all other network connections.
For connection problems see