J22 TDLAS Gas Analyzer
Operating Instructions
Hauser
63
Parameter overview with brief description
Parameter
Prerequisite
Description
User entry
Factory setting
Dew point
method 1
—
Sets the method used to calculate
dew point temperature.
•
Off
•
ASTM
•
ASTM
•
ISO
•
AB
ASTM2
Dew point
method 2
—
Sets the method used to calculate
dew point temperature.
•
Off
•
ASTM
1
•
ASTM
2
•
ISO
3
•
AB
Off
Conversion
type
Used if Dew point is enabled
by selecting a method above.
Sets conversion type used to
calculate dew point temperature.
•
Ideal
•
Real
Ideal
Pipeline
pressure mode
Used if Dew point is enabled
by selecting a method above.
Sets the method the pipeline
pressure is input for the dew point
calculation.
•
Current input
1 to n
•
Fixed value
•
External
value
Fixed value
Pipeline
pressure fixed
Used if fixed value is selected
from Pipeline pressure mode.
Sets a fixed pressure at which the
dew point temperature is
calculated.
Floating point
number
•
50000 mbar a
•
725 psi a
Pipeline
pressure
Used if Current input or
External value is selected
from Pipeline pressure mode.
The pipeline pressure value used
by the dew point calculation based
on the Pipeline pressure mode
setting. Current input is the value
from the selected I/O slot 1 to n.
External value is the value set from
the Modbus fieldbus. See
for more info.
None, read only
None, read only
7.8
Setting peak tracking
The peak tracking submenu controls the software utility that keeps the laser scan centered on the absorption peak.
Under some circumstances, the peak tracking function can get lost and lock onto the wrong peak. If the system alarm
displays, the peak tracking function should be reset.
Navigation
Setup menu → Peak Tracking
▸
Peak tracking
Peak track analyzer control
Peak track reset
Peak track average number
Parameter overview with brief description
Parameter
Prerequisite
Description
User
entry
Factory
setting
Peak track
analyzer control
—
Sets peak tracking capability to OFF or ON.
•
Off
•
On
Off
Peak track reset
Used if peak tracking is set above.
Resets peak tracking.
•
Off
Off
1
ASTM D1142 equation 1
2
ASTM D1142 equation 2
3
ISO 18453 Natural Gas