Electrical connection
Proline Promass Q 500 PROFIBUS PA
36
Hauser
Standard cable
4, 6, 8 cores (2, 3, 4 pairs); pair-stranded with common shield
Shielding
Tin-plated copper-braid, optical cover ≥ 85 %
Capacitance C
Maximum 760 nF IIC, maximum 4.2 µF IIB
Inductance L
Maximum 26 µH IIC, maximum 104 µH IIB
Inductance/resistance
ratio (L/R)
Maximum 8.9 µH/Ω IIC, maximum 35.6 µH/Ω IIB (e.g. in accordance with IEC
60079-25)
Loop resistance
Power supply line (+, –): maximum 5 Ω
Cable length
Maximum 150 m (500 ft), see the following table.
Cross-section
Cable length
Termination
2 x 2 x 0.50 mm
2
(AWG 22)
50 m (165 ft)
2 x 2 x 0.50 mm
2
(AWG 22)
A
B
• +, – = 0.5 mm
2
• A, B = 0.5 mm
2
3 x 2 x 0.50 mm
2
(AWG 22)
100 m (330 ft)
3 x 2 x 0.50 mm
2
(AWG 22)
A
B
• +, – = 1.0 mm
2
• A, B = 0.5 mm
2
4 x 2 x 0.50 mm
2
(AWG 22)
150 m (500 ft)
4 x 2 x 0.50 mm
2
(AWG 22)
A
B
• +, – = 1.5 mm
2
• A, B = 0.5 mm
2
Optionally available connecting cable
Connecting cable for
Ex Zone 1, Class I, Division 1, IIC, IIB
Standard cable
2 × 2 × 0.5 mm
2
(AWG 20) PVC cable with common shield (2 pairs, pair-
stranded)
Flame resistance
According to DIN EN 60332-1-2
Oil-resistance
According to DIN EN 60811-2-1
Shielding
Tin-plated copper-braid, optical cover ≥ 85 %
Содержание Proline Promass Q 500
Страница 279: ......
Страница 280: ...www addresses endress com...